5
Section 1 Specifications
Specifications are subject to change without notice.
Measuring Technique
Infrared Duo scattered light technique for color-independent turbidity measurement
Turbidity in accordance with DIN EN 27027 / TS equivalent DIN 38414
Measuring Range
t-line turbidity: 0.001–4000 FNU/NTU
ts-line, inline turbidity: 0.001–4000 FNU/NTU; TSS content: 0.001 mg/L–50 g/L
hs-line, highline turbidity: 0.001–4000 FNU/NTU; TSS content: 0.001 mg/L–500 g/L TSS
Reproducibility Turbidity < 1 %, Total Suspended Solids (TSS) < 3 %
Measuring Accuracy
Turbidity up to 1000 FNU/NTU:
without calibration < 5 % of the measured value ± 0.01 FNU/NTU
with calibration < 1 % of the measured value ± 0.01 FNU/NTU
Technique variation
coefficient
1 % in accordance with DIN 38402
Response Time 1s <
T90 < 300 s (adjustable)
Calibration Zero point permanently set from the factory, gradient once for the TS content
Cable Length 10 m (33 ft), max. 100 m (328 ft) with extension cable
Ambient Temperature 0 to +40 °C (32 to 104 °F)
Pressure Range
Stainless steel: 6 bar or 60 m (87 psi)
PVC: 1 bar or 10 m (14,5 psi)
Flow Velocity Max. 3 m/s (the presence of air bubbles affects the measurement)
Materials
Optics carrier and sleeve: stainless steel 1.4571 or PVC black
Wiper shaft: stainless steel 1.4104
Wiper arm: stainless steel 1.4581
Wiper rubber: silicone rubber (standard) Optional: Viton
1
(LZX578)
Windows and light guide: quartz glass
O-rings (optics carrier, wiper, windows): NBR (acrylonitrile butadiene rubber)
Housing seals: NBR 70
Sensor connecting cable (hard-wired): 1 cable pair AWG 22 / 12 V DC twisted, 1 cable pair AWG
24 / data twisted, common cable screen, Semoflex (PUR)
Sensor connection plug (hard-wired): type M12 enclosure rating IP 67
Threaded cable fitting: stainless steel 1.4305
Inspection interval On request 1/year service contract with guarantee extension to 5 years
Dimensions
Tank probe: D x L 60 mm x 200 mm (2 x 8 in.)
Probe for pipe installation: D x L 60 mm x 315 mm (2 x 12.4 in.)
(Pipe installation fitting: DN 65 / PN 16 DIN 2633; < 5 bar (73 psi); for pipes from DN 80)
Distance sensor - wall (floor): TS > 10 cm (4 in.), turbidity > 50 cm (20 in.)
Weight
Tank probe: approx. 1.8 kg (63 oz) (t-line: approx. 0.6 kg (21 oz))
Probe for pipe installation: approx. 2.4 kg (85 oz)
Pipe installation fitting: approx. 2.7 kg (95 oz) (without probe)
Pipe installation safety fitting:approx. 18 kg (40 lb) (without probe)
User Maintenance 1 h / month, typical
Certifications CE
1
Viton
®
is a registered trademark of E.I. DuPont de Nemours + Co.