DOC023.52.03232.Dez09SOLITAX scUser Manual12/2009, Edition 4A© HACH-LANGE GmbH, 2006–2009. All rights reserved. Printed in Germany.
10Installation3.2 Sensor InstallationFigure 4 on page 10 illustrates the installation overview for Solitax sc Models t-line, ts-line, and hs-line for
11Installation3.3 Pipe Installation• Install the sensor in an up-flow pipe section for best results. Do not mount the sensor in a down-flow pipe secti
12InstallationFigure 5 Proper Positioning for Insertion into Pipe Direction of Flow
13Installation3.4 Connecting sensor cableCAUTION Always lay cables and hoses so they do not pose a trip hazard and are not bent.1. Unscrew the protect
14Installation
15Section 4 Operations4.1 Use of an sc controllerBefore using the sensor in combination with an sc controller, refer to the controller user manual for
16Operations4.6 CalibrationThere are two calibration techniques; depending on whether turbidity or suspended solid is required (refer to section 4.6.2
17Operations4.6.1 Setting the Outmode1. From the Main Menu, select SENSOR SETUP and press confirm. 2. Select the appropriate sensor if more than one i
18Operations4.6.3 Calibration for Suspended SolidsSuspended solids calibration requires calibration to the actual sample. This optimizes the compensat
19Section 5 MaintenanceDANGEROnly qualified personnel should conduct the tasks described in this section of the manual. Proper maintenance of the meas
20Maintenance5.3 Replacing the WiperThe life of the wiper is dependent on the number of cleaning actions performed and the type of deposits to be remo
21Section 6 Troubleshooting6.1 Error CodesIn the case of an error, the indication of the measured value flashes on the display and all the contacts an
22Troubleshooting
23Section 7 Replacement Parts and Accessories 7.1 Immersion Sensors11 All sensors come with the sensor, replacement wipers, and manual. Description
24Replacement Parts and AccessoriesExtension pipe, 1,0 m (3.28 ft) LZY413 Extension pipe, 1,8 m (5.90 ft) LZY414Installation kit, fixed-point (for t
25Section 8 Contact Information HACH CompanyWorld HeadquartersP.O. Box 389Loveland, Colorado80539-0389 U.S.A.Tel (800) 227-HACH(800) -227-4224(U.S.A.
26Contact InformationHACH LANGE D.O.O.Fajfarjeva 15SI-1230 DomžaleTel. +386 (0)59 051 000Fax +386 (0)59 051 [email protected]ΗΑCH
27Section 9 Warranty and liabilityThe manufacturer warrants that the product supplied is free of material and manufacturing defects and undertakes the
28Warranty and liability
29Appendix A Modbus Register Information Table 3 Sensor Modbus RegistersGroup Name Tag Name Register Data Type# Length R/W DescriptionMeasurements Tu
3Table of ContentsSection 1 Specifications...
30Modbus Register Information
4Table of Contents
5Section 1 SpecificationsSpecifications are subject to change without notice. Measuring TechniqueInfrared Duo scattered light technique for color-inde
6SpecificationsFigure 1 Sensor Dimensions1. SOLITAX sc models t-line, ts-line, and hs-line for immersion in open tanks2. SOLITAX sc models inline and
7Section 2 General Information2.1 Safety InformationPlease read this entire manual before unpacking, setting up, or operating this equipment. Pay atte
8General Information2.2 Sensor OverviewNote: All sensors are also available without wipers for special applications. t-line: 0.001–4000 FNU/NTUHigh-re
9Section 3 InstallationDANGEROnly qualified personnel should conduct the tasks described in this section of the manual. 3.1 Unpacking the InstrumentFi