Hach-lange DR 2800 eco User Manual

Browse online or download User Manual for Equipment Hach-lange DR 2800 eco. Hach-Lange DR 2800 eco User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 130
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - DR 2800 eco

DOC027.53.00791.MRZ07DR 2800DR 2800 eco

Page 2

8Specifications

Page 3

98Advanced Operations• Markierung: to mark the next data point collected. This mark is not used by the instrument, but is available for the user, and

Page 4

99Advanced Operations6.5.3.1 Navigating in the graph of a time scan or a time scan analysisAfter a time scan has been carried out, the time and the ab

Page 5 - Table of Contents

100Advanced Operations6.6 System Check1. Press System Checks in the ”Main Menu“.The ”System Check“ menu contains instrument information and various pe

Page 6

101Advanced Operations6.6.2 Upgrading the Instrument SoftwareYou can obtain the software for the update from the Internet underwww.hach-lange.com >

Page 7

102Advanced Operations6.6.4 Output ChecksIf a printer is connected a test printing of the current screen will be printed.6.6.5 Lamp HistoryThe menu La

Page 8

103Advanced Operations6.7 Favorite Programs (DR 2800 only)The most frequently used tests/methods in the Stored Programs menu and the User Programs men

Page 10 - Specifications

105Section 7 MaintenanceDANGEROnly qualified personnel should conduct the tasks described in this section of the manual.7.1 Cleaning the instrument an

Page 11 - Section 2 General Information

106Maintenance7.1.3.2 Pour-Thru CellImportant Note: Do not use solvents (e.g. acetone) to clean the Pour-Thru Cell. A dilute acid solution can be used

Page 12 - 2.3 Fields of application

107Maintenance7.2 Inserting or changing the battery (DR 2800 only)1. Remove the cuvette from the cell compartment.2. Switch the instrument off.3. Unpl

Page 13 - 2.4.3 Source Lamp Safety

9Section 2 General Information2.1 Safety InformationPlease read this entire manual before unpacking, setting up, or operating this equipment. Pay atte

Page 14

108Maintenance8. Push in the battery so that the contacts of the instrument can be plugged into the socket of the battery. Important Note: Take care t

Page 15 - Section 3 Installation

109MaintenanceCAUTIONPay special attention to the following information for the Lithium battery pack:• Do not short-circuit the contacts.• Keep the ba

Page 16 - 3.4 Front- and back view

110Maintenance7.2.1.4 Loading the battery/Operating timeThe time required to charge a newly installed battery is about 3.5 hours.Operating time of a f

Page 17 - 3.5 Display

111Maintenance7.3 Changing of lamp1. Remove the cuvette from the cell compartment.2. Switch the instrument off.3. Unplug the power cord.CAUTIONIf the

Page 18 - 3.6 Interfaces

112MaintenanceCAUTIONWait until the lamp cools down. Contact with the hot lamp can cause burns.9. Lift the lamp fitting carefully.10. Carefully remove

Page 19 - Installation

113Section 8 TroubleshootingProblem/Display screen Likely Cause ActionAbsorbance > 3.5! The measured absorbance exceeds 3.5Dilute the sample and re

Page 20

114Troubleshooting

Page 21

115Section 9 Replacement Parts and Accessories9.1 Replacements Parts (DR 2800)9.2 Replacements Parts (DR 2800 eco)Description Cat. No.Rechargeable lit

Page 22

116Replacement Parts and Accessories9.3 Accessories (DR 2800)9.4 Accessories (DR 2800 eco)Description Cat. No.USB-Memory Stick LZV568USB-Interface Cab

Page 23

117Section 10 How To OrderOrders/Repair servicePlease contact your representative:Information Required• Hach account number (if available)• Billing ad

Page 24

10General Information2.1.4 General Product InformationThe spectrophotometer has been designed in accordance with the current state of the technology a

Page 25

118How To Order

Page 26 - 3.9 Beam path

119Section 11 Warranty, liability and complaintsThe manufacturer warrants that the product supplied is free of material and manufacturing defects and

Page 27 - Section 4 Start Up

120Warranty, liability and complaints

Page 28 - 4.3 Self Check

121Appendix A Pour-Thru Cell (DR 2800 only)CAUTIONDo not use the Pour-Thru Cell in tests that call for the use of organic solvents such as toluene, ch

Page 29 - Section 5 Standard Operations

122A.2 Assembling the Pour-Thru moduleInstalling Pour-Thru Cell1. Open the cell compartment.2. Insert the pour-thru cell adapter in the cell compartme

Page 30 - 5.1.3 Main Menu

123Figure 17 Pour-Thru Stand Assembly1 Drain Tube 4 Distance between inflow tube into the cell and outflow tube from the cell into the stand pipe = 5

Page 31 - 5.2 Instrument Setup Mode

124Figure 18 Installation Pour-Thru Cell Kit1 Cell compartment (1) for 13 mm round cuvettes with Barcode6 Stand pipe for outflow liquid2 Pour-Thru Ce

Page 32 - 5.2.2 Sample ID

125A.3 Using the Pour-Thru CellMost HACH methods prescribe the use of round cells.When a Pour-Thru cell is used, the measurement results of these HACH

Page 34

127Appendix B Barcode Programs on DR 2800 ecoBarcode Programs on DR 2800 ecoAcid capacity: LCK362Ammonia: LCK302, LCK303, LCK304, LCK305BOD5: LCK554,

Page 35 - 5.2.6 PC und Drucker

11General Information2.4 Important information about the manualThe manual contains all the information needed to enable the instrument to be operated

Page 37 - 5.2.7 Password

12General Information2.5 Overview of Product/FunctionDR 2800The DR 2800 spectrophotometer is a VIS spectrophotometer with a wavelength range of 340 to

Page 38

13Section 3 InstallationDANGEROnly qualified personnel should conduct the tasks described in this section of the manual.3.1 Unpacking the instrument1.

Page 39

14Installation3.2 Operating EnvironmentThe following conditions are necessary to ensure that the instrument runs smoothly and has a long lifespan.• Ma

Page 40 - 5.3.1 The Data Log

15Installation3.5 DisplayAll current selection and input options, analysis results and calibration curves are shown in the graphic display.The display

Page 41

16Installation3.6 InterfacesThe DR 2800/DR 2800 eco has two USB interfaces as a standard feature. They are located on the back wall of the instrument.

Page 42

17Installation3.7 Cell compartments, Cell adapter, Protective Cover and Light ShieldThe DR 2800 spectrophotometer comes equipped with three adapter.Th

Page 44

18InstallationFigure 4 Cell adapter (DR 2800 only)1 10 mm rectangular cell adapter (A) (Cat. No: LZV583) 3 1 inch round cell adapter (C) (Cat. No: LZ

Page 45

19InstallationFigure 6 Light Shield1 Light Shield (Cat. No: LZV646)1

Page 46

20Installation3.7.1 Installing the cell adapter (DR 2800 only)1. Open the cell compartment.2. Select the correct cuvette/cell adapter for your cuvette

Page 47

21Installation3.7.2 Installing the Light ShieldNote: The DR 2800/DR 2800 eco is supplied with the light shield in place. Any transport securing materi

Page 48 - 5.4.2 Using Program Timers

22Installation3.8 Mobile use of the DR 2800 for field analysis (on-site analysis; DR 2800 only)If the DR 2800 spectrophotometer is used to carry out f

Page 49

23Installation3.8.1 Positioning the Protective Cover (DR 2800 only)1. After you have inserted the zero solution cuvette or the sample cuvette, positio

Page 50

24Installation3.9 Beam pathThe following schematic diagram shows the beam path of the DR 2800/DR 2800 eco:Figure 10 Beam path1 Tungsten lamp 8 Filter

Page 51 - 5.4.6 Running a Reagent Blank

25Section 4 Start Up4.1 Switching the Instrument on or off1. Plug in to the power supply (laboratory analysis) or insert the battery (field analysis).

Page 52 - 5.4.7 Analysis of Samples

26Start Up4.3 Self CheckEach time the instrument is powered up, a series of diagnostic tests is performed automatically to ensure operation of major s

Page 53

27Section 5 Standard Operations5.1 Getting Started5.1.1 Tips for Using the Touch ScreenThe entire screen is touch-activated (”Touch Screen”). Spend a

Page 54

DOC027.53.00791.MRZ07DR 2800DR 2800 ecoUSER MANUAL03/2007 edition 1©Hach Lange GmbH, 2007. All rights reserved. Printed in Germany.sd/ck 03/07 edition

Page 55 - 5.5 Barcode Programs

28Standard Operations5.1.3 Main MenuA variety of modes may be selected from the ”Main Menu“. The following table briefly describes each menu option.Le

Page 56

29Standard Operations5.2 Instrument Setup Mode1. Select Instrument Setup in the ”Main Menu“.A selection of functions appears in order to configure the

Page 57

30Standard Operations5. The display shows the chosen Operator ID.6. Press OK. The instrument will return to the ”Instrument Setup“ screen and show the

Page 58

31Standard Operations5. If you want to number the Sample-IDs sequentially (e.g. Inflow (01... etc.)), activate Add Number. • Use the arrow keys to spe

Page 59 - 5.5.5 Sample blank

32Standard Operations5.2.4 Display and Sound Preferences 1. Press Display & Sound on the ”Instrument Setup“ screen. Four options will appear:• Dis

Page 60

33Standard Operations3. Select one of the options under Auto off to set the length of the idle period that can elapse before the instrument automatica

Page 61

34Standard Operations5.2.6.1 Printer SetupNote: For reasons of compatibility, the printer language must be HP PCL 3.2. Press Printer.3. Press Setup to

Page 62

35Standard Operations6. Press Paper to select the paper size.You can choose between• Letter• Legal• Executive and• A47. Press OK to confirm.Note: Pres

Page 63

36Standard Operations3. Use the alphanumeric keypad to enter a new Password (up to 10 characters each) and confirm with OK.The access to the Security

Page 64 - (DR 2800 only)

37Standard Operations5.2.7.1 Deactivation Password1. Press Password in the ”Instrument Setup“ menu.2. Use the alphanumeric keypad to enter the former

Page 66

38Standard Operations5.3 Storing, Recalling, Sending and Deleting Data 5.3.1 The Data LogThe DR 2800 Data Log will store up to 500 readings taken in t

Page 67 - Section 6 Advanced Operations

39Standard Operations4. The function Filter Settings is used to search for specific items.5. Activate On to turn on the filters. To select data by• Sa

Page 68

40Standard Operations5.3.1.3 Sending Data from the Data Log Sending data to a USB memory stick (USB Type A).1. Press Options in the ”Recall Data“ menu

Page 69

41Standard Operations5.3.2 Time Course (DR 2800 only)The instrument can store 20 Time Course Data sets. The data can be stored manually at the user&ap

Page 70

42Standard Operations2. Press Graph to look at details.Note: Press View Summary to return to the Recall Data list.3. Press Table to look at details.No

Page 71

43Standard OperationsOption 2:1. Press Time Course and then Options, More... and Send Data to send the data to a USB memory stick or to a printer.The

Page 72

44Standard Operations5.4 Stored Programs (DR 2800 only)The DR 2800 spectrophotometer contains more than 200 programmed procedures that you can access

Page 73

45Standard OperationsSend Data icon / Send DataTo send Data to a printer, computer or USB memory stick (Type A)Timer iconThis functions as a stopwatch

Page 74

46Standard Operations5.4.2 Using Program TimersSome procedures do not require the use of timers. Other procedures require several timers. These timers

Page 75

47Standard Operations5.4.3 Setting the Dilution FactorYou can use the Dilution Factor function to calculate the original concentration of a sample tha

Page 76

3Table of ContentsSection 1 Specifications ...

Page 77

48Standard Operations5. Press Adjust to enable the Standard Adjust. The Standard Adjust icon will appear.Note: The adjustment must be within certain l

Page 78

49Standard Operations5.4.6 Running a Reagent BlankSome of the stored tests/methods include the "Reagent Blank" function. This enables the re

Page 79

50Standard Operations5.4.7 Analysis of Samples1. Press Stored Programs and select a program.2. Insert the blank cuvette into the cell holder.3. Press

Page 80

51Standard Operations1. Press Stored Programs in the ”Main Menu“ and select a program. 2. Press Program Options and then Edit.3. The Data list of the

Page 81

52Standard Operations5.4.9 Adding Stored Programs to the ”Favorite Programs“ ListCreating a selection list of the most frequently used tests and metho

Page 82

53Standard Operations5.5 Barcode ProgramsDR 2800/DR 2800 ecoA special barcode reader in the cell compartment automatically reads the coded information

Page 83

54Standard Operations5.5.1 Carrying out a LANGE Barcode Cuvette TestImportant Note: Before you carry out a LANGE Barcode Cuvette Test you should inser

Page 84

55Standard Operations5.5.3 Selecting the chemical evaluation formThe form of the chemical evaluation of a number of LANGE test parameters can be selec

Page 85

56Standard Operations5.5.4 Basic test-specific and sample-specific data settings1. Press Options if you wish to change test-specific or sample-specifi

Page 86

57Standard Operations5.5.5 Sample blankTurbidity and color in the sample matrix can falsify the results of a photometric analysis. The interference fa

Page 87

4Table of Contents5.3 Storing, Recalling, Sending and Deleting Data ...

Page 88

58Standard OperationsEnter the namePlease check the working procedure in advance to determine whether this point really needs to be changed.1. Activat

Page 89

59Standard OperationsEnter the Formula, Concentration Units, Designation, Measuring RangesDetailed information for entering the Formula, see 6.1.3 &qu

Page 90

60Standard Operations5.5.6.3 Updating an existing Barcode Cuvette TestNote: Only in very few cases does the revision of a test require all test specif

Page 91

61Standard Operations5.5.6.4 Programming of a new LANGE Cuvette Test (DR 2800 only)1. Insert the Light Shield.2. The instrument need to be at the ”Mai

Page 92

62Standard Operations5.5.7 Updating from the InternetYou can obtain the software for the update from the Internet underwww.hach-lange.com > Intern

Page 93

63Standard Operations5.5.8.1 Evaluation of barcode-readable appliaction with LCW 9061. Insert the Light Shield.2. The instrument need to be at Barcode

Page 94

64Standard OperationsDelete:1. Press Stored Programs in the ”Main Menu“ and select a program. 2. Press Program Options and then Delete.3. Press OK.

Page 95

65Section 6 Advanced Operations6.1 User Programs (DR 2800 only)User programs provide the opportunity of carrying out "made to measure" analy

Page 96

66Advanced Operations6.1.1 Programming a user method1. Select New in the ”Program Options“ menu.All input steps and their significance and options are

Page 97

67Advanced OperationsProgram Type:6. Select the required option and press Next.6.1.2 Single Wavelength and Multi Wavelength programsIf the Single Wave

Page 98

5Table of ContentsSection 6 Advanced Operations ...

Page 99

68Advanced OperationsAbsorbance Formula (Multi-wavelength program type): The Absorbance Formula menu is used to define the wavelengths and the coeffic

Page 100 - Advanced Operations

69Advanced OperationsConcentration resolution (number of decimal places)Select the required number of decimal places from the displayed list and press

Page 101

70Advanced OperationsCalibrationA method is calibrated by determining the absorbance values of several standard solutions of known concentration.There

Page 102 - 6.6 System Check

71Advanced Operations4. When the data have all been entered, you can display the curve that results from plotting the entered data by pressing Graph.

Page 103 - 6.6.3 Optical Checks

72Advanced Operations3. Press the "+" symbol again (see arrow) and enter the next standard concentration. Repeat this sequence until all sta

Page 104 - 6.6.5 Lamp History

73Advanced OperationsEnter Formula:1. Press Enter Formula and press Next.2. Press the formula key. A list of available formulas (linear and 2nd and 3r

Page 105

74Advanced OperationsUpper and lower limits of the measuring rangeIt is possible to enter a maximum (upper) and minimum (lower) measurement value. An

Page 106

75Advanced Operations6.1.3 "Free programming" program typeAn overview of the specifications of the programmed test is displayed. The input o

Page 107 - Section 7 Maintenance

76Advanced Operations6.1.3.2 Entering a new element of a measuring sequence1. Activate the Measurement Process line in the data overview and press Edi

Page 108 - Maintenance

77Advanced Operations[R] key / Reading...1. Press the [R] key to program a measurement measurement of the substance that is to be analyzed. Confirm wi

Page 109

6Table of ContentsSection 8 Troubleshooting ...

Page 110

78Advanced Operations6.1.3.3 Calibration formula (evaluation formula)The calibration formula (evaluation formula) defines the calculation and display

Page 111

79Advanced OperationsContent and use of the keys:Abs/Variables key:1. Press the Abs/Variables key to select, from the displayed list, the required ele

Page 112

80Advanced OperationsThe following basic mathematical operations are available:• + (Addition)• - (Subtraction)• / (Division)• * (Multiplication)• ^ (E

Page 113 - 7.3 Changing of lamp

81Advanced OperationsEntering the next calibration formula C2 or Cn1. Activate C2: Off and press Edit.2. Select C2: Off again and press Edit. The disp

Page 114

82Advanced Operations6.1.3.5 Saving the freely programmed user program1. Press Store to save the entered data.Note: The data can be stored under any d

Page 115 - Section 8 Troubleshooting

83Advanced Operations6.1.5.1 Add to Favorites1. Press Add to Favorites and confirm with OK.The program is added to the Favorites.6.1.5.2 Edit1. Press

Page 116 - Troubleshooting

84Advanced Operations6.2 Standard Addition – Monitoring/Checking Results (DR 2800 only)The accuracy of measured values (their correspondence with the

Page 117

85Advanced Operationsg. Is the pipette volume correct?2. Check the reagents you used, by repeating the standard addition procedure with freshly prepar

Page 118 - 9.4 Accessories (DR 2800 eco)

86Advanced Operations4. Press Options, More... and Standard Addition.An overview of the data of the standard addition procedure is displayed.5. Press

Page 119 - Section 10 How To Order

87Advanced Operations7. Press the keys to enter the standard addition volumes. Use the alphanumeric keypad to enter the new data. Confirm your input b

Page 120 - How To Order

7Section 1 SpecificationsSpecifications are subject to change without notice.Performance SpecificationsOperating Mode Transmittance (%), Absorbance an

Page 121

88Advanced Operations12. Press Ideal line to display the relationship between the added standard solutions and the ideal line (detection rate 100%).

Page 122

89Advanced Operations6.3 Single Wavelength (absorbance, concentration and transmittance measurements; DR 2800 only)The Single Wavelength Mode can be u

Page 123 - A.1.1 Unpacking

90Advanced OperationsConcentration Factor:1. Press Concentration Factor: Off in the ”Options“ menu. Press On to activate this feature.2. Press the ”Fa

Page 124

91Advanced Operations6.3.2 Taking Single Wavelength Measurements1. Insert the blank cuvette into the cell holder. Press Zero.Note:The Read key is only

Page 125

92Advanced Operations6.4 Multi-wavelength mode – measurements at more than one wavelength (DR 2800 only)If measurements are to be carried out at more

Page 126

93Advanced Operationsλ / Absorbance Formula:1. Press Absorbance Formula. 2. The formula selected in the top key determines the number of wavelength an

Page 127 - A.3 Using the Pour-Thru Cell

94Advanced OperationsA 1 refers to the absorbance at wavelength 1A 2 refers to the absorbance at wavelength 2, etc. K 1 refers to the coefficient at w

Page 128

95Advanced Operations6.4.2 Carrying out a measurement in the Multi Wavelength mode1. Insert the blank cuvette into the cell holder. Press Zero. Note:

Page 129

96Advanced Operations6.5 Time Course of absorbance/transmittance (DR 2800 only)The Time Course Mode is used to collect data in either absorbance or tr

Page 130

97Advanced OperationsTime & Interval:1. Press Time & Interval in the ”Options“ menu.2. Input the total time and the reading time and confirm w

Related models: DR 2800

Comments to this Manuals

No comments