Page 1 - PORTABLE TURBIDIMETER

DOC022.52.00630JUN04PORTABLE TURBIDIMETERModel 2100P ISOInstrument manual© Hach Company, 2004. All rights reserved. Original: bb 6/04 4ed

Page 2

Page 10SPECIFICATIONS4650098 specs.fmSPECIFICATIONS, continuedStorage Temperature: –40 to 60 °C (instrument only)Operating Temperature: 0 to 50 °C (in

Page 3 - TABLE OF CONTENTS

Page 114650098 opstop.fmDANGERHandling chemical samples, standards, and reagents can be dangerous. Review the necessary Material Safety Data Sheets an

Page 5 - CERTIFICATION

Page 134650098 intro.fm DescriptionSection 1 Description 1.1 General DescriptionThe Hach Model 2100P ISO Portable Turbidimeter (Figure 1) measures tur

Page 6 - Emissions

Page 14Preparation for Use4650098 intro.fmSection 1, continuedratio of the signals from the 90° and transmitted light detector. This ratio technique c

Page 7 - Certification, continued

Section 1, continuedPage 154650098 intro.fm Preparation for UseFigure 3 Battery Installation1.4.2 Using the Battery Eliminator and Rechargeable Batter

Page 8 - SAFETY PRECAUTIONS

Page 16Preparation for Use4650098 intro.fmSection 1, continued

Page 9 - SPECIFICATIONS

Page 174650098 operation.fm OPERATIONSection 2 OPERATION2.1 Operational Controls and IndicatorsKey DescriptionPower key turns instrument on and off. I

Page 10 - SPECIFICATIONS, continued

Page 18Signal Averaging4650098 operation.fmSection 2, continued2.2 Signal Averaging Signal averaging measures and averages ten measurements while disp

Page 11 - OPERATION

Section 2, continuedPage 194650098 operation.fm Calibrationduring measurement, the initial value is displayed in 12 seconds and is updated every 1.2 s

Page 12 - 4650098 opstop.fm

http://www.hach.com

Page 13 - Section 1 Description

Page 20Calibration4650098 operation.fmSection 2, continued• Store away from direct sunlight. Store vials in their respective kit or shipping box with

Page 14 - 1.4 Preparation for Use

Section 2, continuedPage 214650098 operation.fm Calibration2.3.2 Correcting for Turbidity of Dilution WaterThe 2100P ISO Turbidimeter automatically co

Page 15 - Section 1, continued

Page 22Calibration4650098 operation.fmSection 2, continued• To review a calibration, press CAL and then × to view the calibration standard values. As

Page 16

Section 2, continuedPage 234650098 operation.fm Calibration7. The display will show “S1” (with the 1 flashing) and 20 FNU or the value of the S1 stand

Page 17 - Section 2 OPERATION

Page 24Calibration4650098 operation.fmSection 2, continued17. Press READ. The instrument will count from 60 to 0 (67 to 0 if signal average is on). Th

Page 18 - 2.2 Signal Averaging

Page 254650098 measure.fm TURBIDITY MEASURESection 3 TURBIDITY MEASURE3.1 Turbidity MeasurementMeasurements may be made with the signal average mode o

Page 19 - 2.3 Calibration

Page 26Turbidity Measurement4650098 measure.fmSection 3, continued5. Wipe off the excess so that only a thin coat of oil is left. Excess oil may retai

Page 20 - Section 2, continued

Page 274650098 mainstop.fm MAINTENANCESome of the following manual sections contain information in the form of warnings, cautions and notes that requi

Page 21

Page 28MAINTENANCE4650098 mainstop.fm

Page 22

Page 294650098 maintenance.fm MAINTENANCESection 4 MAINTENANCE 4.1 Lamp Replacement The procedure below explains lamp installation and electrical conn

Page 23

Page 34650098 GBTOC.fmTABLE OF CONTENTSCERTIFICATION ...

Page 24

Page 30TROUBLE SHOOTING4650098 diagnostics.fmSection 5 TROUBLE SHOOTING5.1 Instrument Diagnostics5.1.1 Display TestPressing and holding the I/O key tu

Page 25 - Section 3 TURBIDITY MEASURE

Section 5, continuedPage 314650098 diagnostics.fm Error MessagesMessage Probable Cause Corrective ActionE1Dilution water is ≥ 0,5 FNU. Start calibrati

Page 26 - Section 3, continued

Page 32Replacement Parts & Accessories4650098 parts.fmReplacement Parts & AccessoriesREPLACEMENT PARTSDescription Cat. No.AA Batteries, 4/pkg

Page 27 - MAINTENANCE

, continuedPage 334650098 parts.fm Replacement Parts & AccessoriesStablCal® Calibration Set for 2100P ISO Turbidimeter<0.1, 20, 100, 800 FNU, 5

Page 28

Page 34HOW TO ORDER4650098 end.fmHOW TO ORDERInformation RequiredIn Europe, the Middle East, or Mediterranean Africa:Dr. Bruno Lange GmbH & Co KGW

Page 29 - Section 4 MAINTENANCE

Page 354650098 end.fm REPAIR SERVICEREPAIR SERVICEAuthorization must be obtained from Hach Company before sending any items for repair. Please contact

Page 30 - Section 5 TROUBLE SHOOTING

Page 36WARRANTY4650098 end.fmWARRANTYHach Company warrants this product to the original purchaser against any defects that are due to faulty material

Page 31 - Section 5, continued

, continuedPage 374650098 end.fm WARRANTYThis warranty contains the sole express warranty made by Hach Company in connection with its products. All i

Page 32 - REPLACEMENT PARTS

Page 44650098 GBTOC.fmTable of contents, continuedSection 5 TROUBLE SHOOTING ...30

Page 33 - , continued

Page 54650098 cert.fmCERTIFICATIONHach Company certifies this instrument was tested thoroughly, inspected and found to meet its published specificatio

Page 34 - HOW TO ORDER

Page 64650098 cert.fmCertification, continuedEmissionsInstrument Tested with the external 230 V, 50 Hz Power Supply:Per 89/336/EEC EMC: EN 61326:1998

Page 35 - REPAIR SERVICE

Certification, continuedPage 74650098 cert.fmThis equipment has been tested and found to comply with the limits for a Class A digital device, pursuant

Page 36 - WARRANTY

Page 8SAFETY PRECAUTIONS4650098 safety.fmSAFETY PRECAUTIONSPlease read this entire manual before unpacking, setting up, or operating this instru-ment.

Page 37

Page 94650098 specs.fm SPECIFICATIONSSPECIFICATIONSSpecifications are subject to change without notice. Operating specifications applicable at 25 °C u