DOC023.52.93068Polymetron 9526 ConductivityCertification System11/2013, Edition 3User Manual
Figure 1 Front and rear view1 Controller 6 IP 67 socket for electric powersupply11 Sample inlet2 Protective hood 7 Lead seal 12 Sample ou
The product also uses an accurate temperature compensation algorithm taking account of the purewater dissociation and of any components such as NaCl o
Hydraulic connectionsN O T I C E Sockets No. 4, 5 and 6 in Figure 1 on page 8 are all IP 67 therefore it is essential the connectors are tightenedfirm
Figure 2 Installation in-line1 D1 2 D2Installation off-linePlace the sensor in a flow chamber and connect the flow chamber to the "OUT"
Figure 3 Installation off-lineMains connectionW A R N I N G The installation of the instrument should be performed exclusively by personnel specializ
Disassemble the connector by unscrewing the two ends of the connector (Nos. 1 and 6 in Figure 4)from the main body. Pass the power cable through the c
Figure 5 SD card installationStartupMake sure that the flow rate and pressure do not exceed the values in Specifications on page 3.1. Open the valve
Figure 6 Keypad and front panel overview1 Instrument display 5 BACK key. Moves back one level in the menustructure.2 Cover for secure digital m
Figure 7 Example of Main Measurement screen1 Home screen icon 7 Warning status bar2 Sensor name 8 Date3 SD Memory card icon 9 Analog outp
1. From the graphical display screen use the up and down arrow keys to select a graph and pushthe HOME key.2. Select an option:Option DescriptionMEASU
Option DescriptionDISPLAY SETUP Configures the controller display.ADJUST ORDER—View and modify the measurement display order.• SEE CURRENT ORDER—View
Using the secure digital memory (SD) cardAn SD card must be installed in the controller.• The SD card can be used to update software and firmware and
3. Select the time period from which logs are to be saved.Option DescriptionLAST DAY All logs from the last full 24 hours, starting from 12:00 a.m., a
Firmware updates with SD cardsThe latest firmware updates can be placed on an SD card. The SD card can then be used to updatethe controller or device
d. Push enter to return to the MANAGE CONFIGURATION menu.4. To copy settings from the SD card to a controller (or a device connected to it):a. Select
CalibrationAbout sensor calibrationThere are no calibration options available from the menus of the 9526 instrument. All calibrations areperformed fro
K CalculationUse this option to re-calculate the cell constant K value for the sensor under test.1. Push the menu key and select SENSOR SETUP>K CAL
1. Use a clean, soft cloth to remove loose debris from the end of the sensor. Rinse the sensor withclean, warm water.2. Soak the sensor for 2 to 3 min
Option DescriptionSIMULATION (onlydisplays if sensors ormodules are connected)After the SIM VALUE is entered, the controller outputs this value as if
1. Push the menu key and select DIAGNOSTICS.2. Select the device (controller, sensor, network card) with the warning or error and push enter.3. Select
Table of ContentsAdditional information...3Specifications...
Table 3 Warning list for conductivity sensors (continued)Warning Description ResolutionWRNG BLT-IN TEMP TABLE The measured temperature isoutside of t
Replacement parts and accessories (continued)Description Item no.Cable for the 4-20 mA output (20 meters) 08319=A=0020 ¾’’ NPT flow chamber in PP with
30 English
HACH COMPANY World HeadquartersP.O. Box 389, Loveland, CO 80539-0389 U.S.A.Tel. (970) 669-3050(800) 227-4224 (U.S.A. only)Fax (970) 669-2932orders@hac
Operation...22Configure the sensor u
Additional informationAdditional information is available on the manufacturer's website.SpecificationsSpecifications are subject to change withou
Specification DetailsCell constant K 0.01 (cm-1)Conductivity range 0.01—200 μS.cm-1; Resistivity range: 5k Ω.cm—100 MΩ.cmMaximum pressure 10 barMaximu
Specification DetailsData logging Secure Digital Card (32 GB maximum) or special RS232 cable connector for datalogging and performing software updates
Precautionary labelsRead all labels and tags attached to the product. Personal injury or damage to the product couldoccur if not observed. A symbol on