Hach-lange POLYMETRON 9240 Operator Manual User Manual

Browse online or download User Manual for Equipment Hach-lange POLYMETRON 9240 Operator Manual. Hach-Lange POLYMETRON 9240 Operator Manual User Manual

  • Download
  • Add to my manuals
  • Print
  • Page
    / 144
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews

Summary of Contents

Page 1 - Analyzer

DOC024.52.93034POLYMETRON Model 9240 Multi-Channel SodiumAnalyzerUser Manual09/2013, Edition 8

Page 2

8General Information1.4 Product recycling informationENGLISHElectrical equipment marked with this symbol may not be disposed of in European public dis

Page 3 - Table of Contents

98Calibration8.4.3 Two point calibrationWhen prompted, fill the overflow vessel with the calibration solution and select OK to start the measurement.T

Page 4

99Calibration8.5 Automatic calibration setupSelect the AUTO. CAL. SETUP option to access its sub-menu and set the calibration parameters.This allows t

Page 5

100Calibration8.6 Calibration resultsAfter the slope and offset have been calculated, the default parameters, last calibration details and current cal

Page 6

101Calibration8.8 Custom adjustment8.9 Temperature calibrationWhere the analyzer displays values that are slightly above or below the expected value,

Page 8 - 1.3.3 Service and repairs

103Section 9 Maintenance and DiagnosticsPeriodic maintenance will ensure accurate and consistent analysis results. Verify the levels of the calibratio

Page 9 - 1.3.5 Precautionary labels

104Maintenance and Diagnostics9.3 Fuse replacementThe following illustration shows the position of the fuse, which is located next to the mains connec

Page 10 - General Information

105Maintenance and Diagnostics9.5 Maintenance and diagnostics menu optionThere are a number of maintenance and diagnostic procedures that can be perfo

Page 11

106Maintenance and Diagnostics9.7 Reagent changesNote: The conditioning reagent and electrolyte consumption are approximate measurements, so a discrep

Page 12 - 1.5 Product disposal

107Maintenance and Diagnostics9.7.2 Priming tubes9.7.3 Bottles fullAfter refilling the bottles (reference electrolyte, calibration solution and reacti

Page 13

9General InformationSVENSKAElektronikutrustning som är märkt med denna symbol kanske inte kan lämnas in på europeiska offentliga sopstationer efter 20

Page 14

108Maintenance and Diagnostics9.8 Calibration diagnostics9.9 Raw values9.10 Test accessories This option is used by after sales service technicians to

Page 15 - Section 2 Specifications

109Maintenance and Diagnostics9.10.1 Hydraulics9.10.2 RelaysTEST HYDRAULICS allows you to check that all accessories apart from relays and logical inp

Page 16 - Specifications

110Maintenance and Diagnostics9.10.3 Logical inputs9.11 Sensor reactivationChoose this option to check the logical inputs status.Each logical input sh

Page 17

111Maintenance and Diagnostics9.12 Extended stopIf the instrument is not to be used for an extended period of time, select this option to shut the ana

Page 18

112Maintenance and Diagnostics9.13 Startup9.14 Software versionsThis process guides you through various steps required to set the instrument up for in

Page 19 - Section 3 Analyzer Overview

113Section 10 Troubleshooting10.1 General faultsWARNINGBefore opening the instrument, ensure the power supply has been switched off.Table 11 General

Page 20 - Analyzer Overview

114TroubleshootingGrab sample is not correct.Calibration is not correct. See below.Activation has been skipped. Start a new grab sample again.Electrod

Page 21 - 3.3.1 Analyzer front panel

115TroubleshootingPump not working(cationic applications).Option not activated. Call a Service Technician.Pump is not connected.Verify the electrical

Page 22 - 3.3.2 Analyzer rear panel

116Troubleshooting10.2 Detection of functional faults10.3 Miscellaneous problemsTable 12 Lack of precisionPROBLEM CAUSE/SOLUTIONElectrolyte is pollut

Page 23 - 3.3.3 Pumps box

117Section 11 Spare Parts and Accessories11.1 Accessories - Options - Maintenance kitsTable 14 Accessories, options & maintenance kitsDescription

Page 24 - 3.4 Conditioning reagent

10General Information1.5 Product disposalNote: The following only applies to European customers.Hach Lange is committed to ensuring that the risk of a

Page 25 - 3.5 pH regulation

118Spare Parts and Accessories11.2 Spare parts - In contact with sample11.3 Spare parts - In contact with cell or electrodesTable 15 Spare parts - In

Page 26

119Spare Parts and Accessories11.4 Spare parts - In contact with reagents or standard solution11.5 ElectronicsTable 17 Spare parts - In contact with

Page 27 - 3.6 Measurement process

120Spare Parts and Accessories11.6 Additional hardwareTable 19 Additional hardwareDescription Part NumberLocking key for 924x enclosure 32965Elbow to

Page 28 - 3.6.2 Measurement steps

121Section 12 Default Configuration12.1 User configuration tableUser ConfigurationOption Default Value Possible Values Selected Values DateMEASUREMENT

Page 29

122Default ConfigurationALARMS / ALARM 3Mode Limit LimitActive channelSample flowNoneAttrib Channel 1 Channel 1 - 4Limit 1.00 ppm 0 - 9990 ppmDirectio

Page 30

123Default ConfigurationmA OUTPUTS / OUTPUT 1Attrib Channel 1 Channel 1 - 4Temper.mVNoneType 0-20 mA 0-20 mA4-20 mAMode Linear LinearDualLogarithmLow

Page 31 - 3.6.4 On-line measurement

124Default ConfigurationmA OUTPUTS / OUTPUT 4Attrib None Channel 1 - 4Temper.mVNoneType 0-20 mA 0-20 mA4-20 mAMode Linear LinearDualLogarithmLow 0 ppb

Page 32

125Default ConfigurationmA OUTPUTS / EVENT INDICATION / WARNING ALARMAttrib None NoneOutput 0 - 5Mode Frozen Preset valueFrozenmA OUTPUTS / EVENT INDI

Page 33

126Default Configuration

Page 34 - 3.8 Manual calibration cycle

127Section 13 Material Safety Data Sheets (MSDS)13.1Diisopropylamine (DIPA)1. IDENTIFICATION OF THE SUBSTANCECatalogue No: 803646Product name: diisopr

Page 35

11General Information1.6 Restriction of hazardous substances (RoHS)The European Union RoHS Directive and subsequent regulations introduced in member s

Page 36 - 3.10 Analyzer outputs

128Material Safety Data Sheets (MSDS)9. PHYSICAL AND CHEMICAL PROPERTIESForm: liquidColor: colorlessOdor: amine-likePh value: not availableMelting tem

Page 37

129Material Safety Data Sheets (MSDS)14. TRANSPORT INFORMATIONDOT: • Shipping Name: Diisopropylamine• UN Number: UN1158• Hazard Class: 3,8• Packing Gr

Page 38 - 3.11 Available options

130Material Safety Data Sheets (MSDS)13.2 Potassium chloride1. IDENTIFICATION OF THE SUBSTANCEIntended/recommended use: electrolyte for reference elec

Page 39 - Section 4 Installation

131Material Safety Data Sheets (MSDS)8. EXPOSURE CONTROLS AND PERSONAL PROTECTIONProtective clothing: None: Overalls: Glasses: Glo

Page 40 - 4.3 Instrument mounting

132Material Safety Data Sheets (MSDS)13.3 Sodium chloride1. IDENTIFICATION OF THE SUBSTANCEIntended/recommended use: 10 ppm (Na) calibration standard

Page 41 - 4.3.2 Wall mounting

133Material Safety Data Sheets (MSDS)8. EXPOSURE CONTROLS AND PERSONAL PROTECTIONProtective clothing: None: X Overalls: Glasses: Gl

Page 42 - 4.4 Step-by-step installation

134Material Safety Data Sheets (MSDS)13.4 Sodium nitrate1. IDENTIFICATION OF THE SUBSTANCEIntended/recommended use: 0.5 M NaNO3 reactivation solution

Page 43

135Material Safety Data Sheets (MSDS)8. EXPOSURE CONTROLS AND PERSONAL PROTECTIONProtective clothing: None: X Overalls: Glasses: Gl

Page 44

136Material Safety Data Sheets (MSDS)

Page 45

137Section 14 Reagent Preparation14.1 Conditioning Reagent• Use of Diisopropylamine in the conditioning bottle is the Hach Lange recommended solution

Page 46

12General Information

Page 47 - 4.4.2 RS485 connection

138Reagent PreparationStandard Solution 100 ppb This solution is of LOW concentration for manual calibration and verification samples.1. Rinse at leas

Page 48

139Reagent Preparation14.4 3M KCl1. To prepare 1L of 3M KCl, rinse at least 3 times a 1L volumetric flask (“A” class) with ultra pure water. The clean

Page 49

140Reagent Preparation

Page 51 - 4.4.5 Reagents installation

HACH COMPANY World HeadquartersP.O. Box 389, Loveland, CO 80539-0389 U.S.A.Tel. (970) 669-3050(800) 227-4224 (U.S.A. only)Fax (970) 669-2932orders@hac

Page 52

13Section 2 Specifications2.1 Technical specificationsSpecifications are subject to change without notice.Table 1 Technical specificationsPERFORMANCE

Page 53

14SpecificationsMECHANICAL SPECIFICATIONSMaximum panel dimension (H x L x D)850 x 450 x 252.5mm [33.46 x 17.71 x 9.94in]InletSimple fittings for 6 mm

Page 54 - 4.4.8 Flow rate adjustment

15Specifications2.2 Model identification systemThe analyzer identification number and the instrument serial number are located on the label on the bac

Page 55

16Specifications

Page 56

17Section 3 Analyzer Overview3.1 OverviewThe Polymetron 9240 Sodium Analyzer is a continuous on-line monitor for direct measurement of sodium in power

Page 58

18Analyzer Overview3.2 Schematic process overviewThe illustration below shows the major components of the analyzer.Figure 2 Working principal1 - Samp

Page 59

19Analyzer Overview3.3 Presentation of the analyzer3.3.1 Analyzer front panelFigure 3 Analyzer front panel1 - User interface 5 - Door lock2 - Overflo

Page 60 - 4.6 Analyzer setup

20Analyzer Overview3.3.2 Analyzer rear panelFigure 4 Analyzer rear panel1 - Local controller box 4 - Sample inlet valves2 - Electrolyte reservoir 5 -

Page 61 - 5.1 Data entry

21Analyzer Overview3.3.3 Pumps boxThree pumps have been factory installed in the pump box and are used for the following purposes:Figure 5 Pumps box

Page 62 - 5.1.2 Numeric fields

22Analyzer Overview3.4 Conditioning reagentHach Lange highly recommends the use of Diisopropylamine (DIPA) as the conditioning reagent. Other reagents

Page 63 - 5.1.3 Alphanumeric fields

23Analyzer Overview3.5 pH regulation3.5.1 Non-cationic applicationsIn order to ensure the accuracy and the repeatability of low sodium concentrations,

Page 64 - 5.1.6 Data entry input errors

24Analyzer OverviewThe usual Tgas/Twater ratio values are as follows:Refer to Sample pH conditioning check on page 53 for the procedure to select the

Page 65 - 5.2 Measurement screens

25Analyzer Overview3.6 Measurement process3.6.1 Smart rinse optionThe measurement cell is rinsed by the analyzer after a calibration, grab sample, or

Page 66 - 5.2.2 Historical display

26Analyzer Overview3.6.2 Measurement stepsThe 9240 is designed as a multi channel analyzer, though it can be used as a single channel analyzer. The me

Page 67 - 5.2.3 Alarms screen

27Analyzer Overview3.6.2.1 Search stability set to NO - Fixed measurement modeAs the cycle and on-line measurement times defined by the user are fixed

Page 68 - 5.3 Main menu

1Table of ContentsSection 1 General Information...

Page 69 - 5.3.1 Verification

28Analyzer Overview3.6.2.2 Search stability set to YES - Automatic measurement modeIn this configuration, a maximum cycle time not to be exceeded has

Page 70

29Analyzer Overview3.6.3 Measurement cycleThe following flow chart illustrates the logic used during the measurement cycle.3.6.4 On-line measurementOn

Page 71 - 5.3.2 Grab sample

30Analyzer Overview3.7 Automatic calibration cycleThe following section gives an overview of the automatic calibration process. For details on setting

Page 72

31Analyzer OverviewThis process calculates the slope and offset of the ISE sodium electrode and the reference electrode. It is based on the measuremen

Page 73 - 5.4 Menu structure overview

32Analyzer Overview3.8 Manual calibration cycleThe following section gives an overview of the manual calibration process. For details on setting up an

Page 74

33Analyzer Overview4. Phase 4: The overflow vessel is drained, and the system waits for the operator to fill the overflow vessel with the low value ca

Page 75 - Section 6 System Setup

34Analyzer Overview3.10 Analyzer outputsThe following tables shows the outputs (screen, analog, alarm relays and RS485) generated by the analyzer duri

Page 76 - 6.2 Date and time

35Analyzer Overview3.10.4 Auto calibration process3.10.5 Manual calibration processAction Screen4-20 mAAlarmsRS485Conc. Other prog. Address 0 OthersPh

Page 77 - 6.3 Display options

36Analyzer Overview3.11 Available options3.11.1 K-Kit (cationic)For a high acidity water such as that from a cation exchanger outlet, the regular gase

Page 78 - 6.5 Default values

37Section 4 Installation4.1 Analyzer inspection and unpackingThe instrument has been factory tested and checked prior to shipping. We nevertheless rec

Page 79 - 6.7 Factory settings

2Table of Contents4.4.3 Input/Output connections ...474

Page 80 - System Setup

38Installation4.3 Instrument mounting4.3.1 Panel mountingFor the panel mount model, the dimensions of the panel and fixation holes are as follows: All

Page 81 - Section 7 User Setup

39Installation4.3.2 Wall mounting Use the wall mounting kit to fix the instrument to the wall.Use these to drill the four holes for fixing the instrum

Page 82 - User Setup

40Installation4.4 Step-by-step installationThe electrical installation should be carried out by duly qualified personnel. A supply voltage of 100-240

Page 83 - 7.2 Measurement

41Installation2. Pass the power supply cable through the cable gland located at the back left of the bottom of the cabinet. 3. Open the I/O connection

Page 84 - 7.2.4 Reactivation frequency

42Installation5. ...and allow the I/O box to rotate down.6. Pass the mains power cable through the back end cable gland on the transmitter enclosure.7

Page 85 - 7.2.5 Datalogger setup

43Installation8. Remove the metallic shielding plate protecting access to the main board.9. Take the power supply connector and note where the earth,

Page 86 - 7.3 Alarms

44Installation11. ...and put the connector back in place.12. Replace the metallic shielding plate.

Page 87 - 7.3.1 Alarms 1 to 4

45Installation4.4.2 RS485 connection1. Connect the RS485 communication cable as indicated. Connection is the same on the CPU board for both the JBUS/M

Page 88 - 7.3.3 System alarm

46Installation3. Put the I/O box back in its normal position.4. Fix the I/O box back in place with the 2 screws.

Page 89

47Installation4.4.3 Input/Output connections1. Analog output – Iout 0 is used for the current signals of the measurement. Other analog outputs can be

Page 90 - 7.4 mA outputs

3Table of Contents7.5 RS485 (or PROFIBUS) ...

Page 91 - 7.4.2 Event indication

48Installation4.4.4 Sample tubes installation• Sample inlet connections - 6 mm (or ¼'') in polyethylene or PTFE or FEP. If particulate matte

Page 92 - 7.5 RS485 (or PROFIBUS)

49Installation4.4.5 Reagents installation1. Prepare the reagents according to Reagent Preparation on page 137. Install the reagent canisters in their

Page 93 - 7.6 Sample channels

50Installation4.4.6 Magnetic stirrer installation1. On the front of the panel, remove the plastic bag from the overflow vessel2. Remove the magnetic s

Page 94 - 7.6.4 Channel names

51Installation4.4.7 Reagents volume declarationNote: As you will now be using the analyzer menus to input data, it may be useful to familiarize yourse

Page 95 - Section 8 Calibration

52Installation4.4.8 Flow rate adjustment1. From the MAINTENANCE/DIAG. menu select START UP and press Enter2. First, the system automatically primes bo

Page 96 - 8.4 Start calibration

53Installation4.4.9 Sample pH conditioning checkNote: At this stage, the electrodes should not have been installed.4.4.9.1 Non-cationic applications4.

Page 97

54Installation7. For each channel, measure the pH in the conditioned sample to check if the pump ratios are efficient enough to obtain a pH of around

Page 98

55Installation5. Install the reference electrode in the extreme left measurement chamber.6. Connect the reference cable (the one without the blue labe

Page 99 - 8.4.2 One point calibration

56Installation4.4.11 Sodium ion selective electrode installationNote: It is critical to preserve the integrity of the sodium ion selective electrode a

Page 100 - 8.4.3 Two point calibration

57Installation4.4.12 Fill electrolyte reservoir1. The electrolyte reservoir is located at the back of the analyzer. 2. Take the KCl electrolyte bottle

Page 101

4Table of ContentsSection 13 Material Safety Data Sheets (MSDS) ...12713.1 Diisopro

Page 102 - 8.7 Calibration loggings

58Installation4.5 Analyzer stabilizationAt this stage the analyzer has been completely installed, but needs to run for a period of time to stabilize.

Page 103 - 8.9 Temperature calibration

59Section 5 Operating InstructionsNote: All screen display examples in the following sections are shown as black on a white background for reasons of

Page 104

60Operating Instructions5.1.2 Numeric fieldsThese fields require that the user enter one or more numeric values into a field. The type of field determ

Page 105 - 9.1 Maintenance schedule

61Operating Instructions5.1.3 Alphanumeric fieldsThese fields require that the user enter one or more alphanumeric values into a field. The type of fi

Page 106 - 9.3 Fuse replacement

62Operating Instructions5.1.4 List element fieldsThis type of data entry is where a pre-defined list of available data values are available to the use

Page 107 - Maintenance and Diagnostics

63Operating Instructions5.2 Measurement screens5.2.1 Principal displayAn example of this display is given above. It shows the details of the current s

Page 108 - 9.7 Reagent changes

64Operating Instructions5.2.2 Historical displayThis screen shows the last sample measurements for each channel, the last grab sample measurement and

Page 109 - 9.7.3 Bottles full

65Operating Instructions5.2.3 Alarms screenAlarms S1-S4 relate to the four alarm outputs. This is followed by the warning alarm (W!) and the system al

Page 110 - 9.10 Test accessories

66Operating Instructions5.3 Main menuFor a diagrammatic view of the complete menu structure, please refer to Menu structure overview on page 71.The ma

Page 111 - 9.10.2 Relays

67Operating Instructions5.3.1 VerificationThis option allows you to verify the measurement using a solution of known sodium concentration.Concentratio

Page 112 - 9.11 Sensor reactivation

5Section 1 General Information1.1 DisclaimerThe information in this manual has been carefully checked and is believed to be accurate. However, Hach La

Page 113 - 9.12 Extended stop

68Operating InstructionsUpon completion of the rinsing step, the analyzer is ready for the first manual step in the cycle. Place the lid of the overfl

Page 114 - 9.14 Software versions

69Operating Instructions5.3.2 Grab sampleNote: It is recommended that the grab sample should have a concentration of > 10 ppb and should be at the

Page 115 - Section 10 Troubleshooting

70Operating InstructionsThe display switches back to the main measurement screen and displays the measurement value.The progress bar at the top of the

Page 116 - Troubleshooting

71Operating Instructions5.4 Menu structure overview

Page 117

72Operating Instructions

Page 118 - 10.3 Miscellaneous problems

73Section 6 System SetupBefore attempting to setup the analyzer, ensure that you have read and understood how to enter and update data fields as descr

Page 119

74System Setup6.1 System setup - Menu overview6.2 Date and timeFigure 20 System setup menuSelect the DATE/TIME option to access the date/time sub-men

Page 120

75System Setup6.3 Display optionsSelect the DISPLAY option to set up the display parameters.Select the LANGUAGE, CONC. UNIT and TEMPERATURE UNIT field

Page 121 - 11.5 Electronics

76System Setup6.4 Passwords6.5 Default valuesSelect the PASSWORDS option to set passwords for access to programming, calibration and system setup opti

Page 122 - 11.6 Additional hardware

77System Setup6.6 Adjust mA output6.7 Factory settingsSelect the ADJUST mA OUTPUT option to access the analyzer’s analog output parameters. These can

Page 123 - 12.1 User configuration table

6General Information1.3.2 Safety recommendationsFor safe operation, it is imperative that these service instructions be read before use and that the s

Page 125

79Section 7 User SetupBefore attempting to setup the analyzer, ensure that you have read and understood how to enter and update data fields as describ

Page 126

80User Setup7.1 User setup - Menu overviewFigure 21 User setup menu

Page 127

81User Setup7.2 Measurement7.2.1 Targeted pH (non-cationic applications only)7.2.2 Total gas/water ratio (cationic applications only)7.2.3 Measure ste

Page 128

82User Setup7.2.4 Reactivation frequencyIf the analyzer has been set up as a multi channel analyzer the screen illustrated left will be displayed.The

Page 129 - 13.1Diisopropylamine (DIPA)

83User Setup7.2.5 Datalogger setupThis option allows you to view data held in the analyzer’s internal memory.Select the VIEW DATA option to display th

Page 130

84User Setup7.2.6 Graph time base7.3 AlarmsFor graphical displays, enter the number of hours as the base line for the graph. This can be 4, 8, 12, 16,

Page 131

85User Setup7.3.1 Alarms 1 to 4Table 4 Alarms 1 to 4 parametersValue DescriptionModeLimitTrigger the alarm when the measurement is above or below a p

Page 132 - 13.2 Potassium chloride

86User Setup7.3.2 Warning alarm7.3.3 System alarmTable 5 Warning alarm parametersValue DescriptionAlarmYes Activate the warning alarmNo Deactivate th

Page 133

87User Setup7.3.4 System and warning alarm tableThe following table lists the different system and warning alarms:Table 7 System and warning alarmsMe

Page 134 - 13.3 Sodium chloride

7General Information1.3.5 Precautionary labelsRead all labels and tags attached to the analyzer. Personal injury or damage to the analyzer could occur

Page 135

88User Setup7.4 mA outputs7.4.1 Outputs 0 to 5Select the mA OUTPUTS option to set up the parameters for all the analog outputs.From the list available

Page 136 - 13.4 Sodium nitrate

89User Setup7.4.2 Event indicationFigure 22 Linear and dual slopesThe screen illustrated left shows an analog output to be activated on a measurement

Page 137

90User Setup7.4.3 Test7.5 RS485 (or PROFIBUS)Note: If the PROFIBUS option has been installed, then the menu option will show PROFIBUS rather than RS48

Page 138

91User Setup7.6 Sample channels7.6.1 Number of channelsEnter the parameters as described in Table 9 on the previous page.Press the Enter function key

Page 139 - 14.2 Standard Solutions

92User Setup7.6.2 Channel activation7.6.3 Sequence7.6.4 Channel namesUsing the Up and Down Arrow keys, define whether the channel is active (Activ) or

Page 140 - Reagent Preparation

93Section 8 Calibration8.1 GeneralBefore attempting to calibrate the analyzer, ensure that you have read and understood how to enter and update data f

Page 141 - 14.5 0.5M NaNO

94Calibration8.3 Calibration - Menu overview8.4 Start calibration8.4.1 Calibrate known additionFigure 23 Calibration menuSelect the START CALIBRATION

Page 142

95CalibrationIf the volume of solution is OK, the background point measurement (P0) to be used as the base measurement value is taken. This is the mea

Page 143

96CalibrationIf the volume of solution is OK, the background point measurement (P0) to be used as the base measurement value is taken. This is the mea

Page 144

97Calibration8.4.2 One point calibrationWhen the measurement is stable, the details are displayed at the top of the screen against measurement P1.The

Comments to this Manuals

No comments