User Manual
DOC022.52.80197
General use pH probe: Models PHC20101, PHC20103
Safety information
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12
August of 2005. In conformity with European local and national regulations (EU Directive 2002/98/EC), European
electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxillary items for proper disposal.
Specifications
Note: Specifications are subject to change without notice.
Specifications Details
Probe type Digital combination non-refillable, semi-liquid gel-filled probe with double
junction reference and built-in temperature sensor
pH range pH 0 to 14
pH resolution User-selectable stabilization time and resolution—Fast: 0.1, Fast: 0.01,
Medium: 0.01, Slow: 0.01 or Slow: 0.001
Slope -59 mV/pH (85 to 115% at 25 °C (77 °F) per Nernstian theoretical value)
Operating temperature range 0 to 80 °C (32 to 176 °F)
Storage temperature range 5 to 40 °C (41 to 104 °F)
Junction Ceramic
Reference type Ag/AgCl
Sodium (alkalinity) error <20% at pH 14 in 1 M NaCl
Temperature accuracy ±0.3 °C (±0.54 °F)
Minimum sample depth 15 mm (0.59 in.)
Dimensions (standard) Diameter: 12 mm (0.47 in.)
Length: 103 mm (4.056 in.)
Total length: 200 mm (7.9 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Cable connection M12 digital output and connector compatible with HQd meters
Product overview
The PHC201 series probe is a non-refillable, semi-liquid gel-filled combination pH probe
with a built-in temperature sensor (Figure 1). The standard PHC20101 or PHC20103 probe
comes with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The probe
provides pH measurement in general aqueous applications. The probe is not suitable for
use with organic solvents, low ionic strength, wastewater, high solids or high alkaline sample
types.
1
Page view 0
1 2 3 4 5 6 ... 13 14

Summary of Contents

Page 1 - User Manual

User ManualDOC022.52.80197General use pH probe: Models PHC20101, PHC20103Safety informationPrecautionary labelsRead all labels and tags attached to th

Page 2 - Calibration

2. Fill the probe storage cap or soaker bottle half full with Hach Electrode Storage Solutionor 3 M potassium chloride (KCl) solution.3. Put the soake

Page 3 - Calibration procedure PHC201

Message or symptom Possible cause ActionCalibration errors Calibration not done correctly Recalibrate using freshly prepared pH buffers.Contaminated g

Page 6 - Change calibration options

HACH COMPANY World HeadquartersP.O. Box 389, Loveland, CO 80539-0389 U.S.A.Tel. (970) 669-3050(800) 227-4224 (U.S.A. only)Fax (970) 669-2932orders@hac

Page 7

Figure 1 Probe overview1 Probe soaker bottle 4 Reference junction2 Standard probe (1 or 3 meter cable) 5 Glass bulb and temperature sensor3 Soaker bo

Page 8 - Change check standard options

Calibration notes• pH buffers can be used in any order. Use buffers that are two pH units apart.• Additional standard sets along with the minimum numb

Page 9 - Maintenance

Sample measurementBefore measurement:The probe must have the correct service-life time stamp. Set the date and time in the meter before the probe is a

Page 10 - Troubleshooting

Run a check standardThe run check standard feature validates instrument performance between samplemeasurements. Use the run check standard feature for

Page 11 - DOC022.52.80197 11

Change measurement optionsMethods are groups of default or user-defined settings relevant to specific applications. Ifthe meter is set to the default

Page 12 - 12 DOC022.52.80197

4. Select Calibration Options and update the settings:Option DescriptionBuffer Set Sets the temperature compensated buffer set used for calibration—•

Page 13

Table 2 Custom buffer setsBuffer set values Option DescriptionStd1Std2Std3Std4Std51.09pH—25 °C Pre-set temperature compensated buffer values.Note: Se

Page 14

5. Select Check Standard Reminder and update the settings:Option DescriptionReminder Repeat Sets the time interval for the check standard reminder—Off